Estimating Mechanical Properties of Bovine Knee Ligaments and Tendons with Near Infrared Spectroscopy

N/A
N/A
Info
Lataa
Protected

Academic year: 2022

Jaa "Estimating Mechanical Properties of Bovine Knee Ligaments and Tendons with Near Infrared Spectroscopy"

Copied!
3
0
0

Kokoteksti

(1)

UEF//eRepository

DSpace https://erepo.uef.fi

Rinnakkaistallenteet Luonnontieteiden ja metsätieteiden tiedekunta

2018

Estimating Mechanical Properties of Bovine Knee Ligaments and Tendons with Near Infrared Spectroscopy

Torniainen, Jari E

The Optical Society

other

info:eu-repo/semantics/acceptedVersion

© Optical society All rights reserved

http://dx.doi.org/10.1364/TRANSLATIONAL.2018.JW3A.29

https://erepo.uef.fi/handle/123456789/7442

Downloaded from University of Eastern Finland's eRepository

(2)

Estimating Mechanical Properties of Bovine Knee Ligaments and Tendons with Near Infrared

Spectroscopy

Jari E. Torniainen, Aapo Ristaniemi, Lauri Stenroth, Juha T¨oyr¨as

Biophysics of Bone and Cartilage Research Group, Department of Applied Physics, University of Eastern Finland jari.torniainen@uef.fi

Abstract: In this study, mechanical properties of bovine knee ligament and tendon samples were estimated using near infrared spectroscopy (NIRS). Properties related to sample stress-relaxation characteristics were found to be suitable for NIRS-based estimation.

OCIS codes: 170.3890, 300.0300

1. Introduction

Arthroscopic evaluation of the connective tissues of the knee (cartilage, meniscus and ligaments) is an important di- agnostic technique for examining knee injuries and providing supplementary information about the tissue condition during repair surgeries. An estimate of the tissue condition is typically obtained through a visual inspection with an arthroscope and manual palpation with a metal hook. The evaluation of the tissue is based on the subjective interpre- tation of the orthopedic surgeon, thus impairing the reliability of the method.

Arthroscopy can be improved by utilizing near infrared spectroscopy (NIRS) for real-time analysis of tissue condi- tion. NIRS is capable of non-destructively determining the chemical composition of biological samples. As mechanical properties of connective tissues correlate with their chemical composition, NIRS can characterize the overall condition and health of the tissue. Effectiveness of NIRS has already been demonstrated for cartilage [1] and meniscus [2] but not for ligaments. Quantified evaluation of ligament condition through NIRS could significantly improve comprehensive diagnostics of joints, such as knee and shoulder.

In this study, we tested NIRS for evaluating mechanical properties of bovine knee ligaments and tendons. NIRS and viscoelastic mechanical properties were measured from 40 samples. Variable selection methods were used to determine which mechanical variables can be reliably predicted with NIRS using partial least-squares regression (PLSR) models.

2. Materials and Methods

The four main ligaments (anterior cruciate, posterior cruciate, lateral collateral and medial collateral) and patellar tendon were extracted from eight bovine knee joints (age 12-24 months). Ligaments and tendons were cut into smaller dumbbell-shaped sample pieces in orientation parallel to the fascicles (12 mm in length, 2 mm wide and thick) using a custom-made punch. Subsequent NIRS and biomechanical measurements were performed with the samples submerged in phosphate buffered saline solution supplemented with enzyme inhibitors.

NIR absorption spectra were measured with a spectrometer (AvaSpec-NIR256-2.5-HSC,λ= 1000-2500 nm, resolu- tion = 6.4 nm, Avantes BV, Apeldoorn, Netherlands) and a custom arthroscopic NIRS probe. Detector integration time was set to 20 ms and the acquired signal was averaged 100 times. Device was calibrated prior to each measurement using a standard reflector (Spectralon SRS-99, Labsphere Inc., North Sutton, USA). Measurements were taken from the surface of the sample at three equispaced locations. Absorption spectra were smoothed using a Savitzky-Golay filter (3rd order polynomial, filter window 39.3 nm) and the second derivate of the smoothed spectrum was used for the statistical analysis.

Viscoelastic properties of the samples were obtained from uniaxial stress-relaxation, sinusoidal and ultimate strength tests. The samples were fixed to a custom mechanical testing apparatus using custom grips. Force and displacement were measured and stress and strain were calculated based on sample dimensions. In total, 49 variables were produced by the mechanical testing. Full description of the mechanical variables is omitted here but an interested reader is directed to [3] for more details.

(3)

3. Statistical Analysis

Statistical analysis aimed to reduce the full set of mechanical variables to a subset which can be reliably predicted using NIRS. This process consisted of three steps: leave-one-out (LOO) analysis, spatial sensitivity analysis and five- fold cross-validation. The LOO analysis rejected mechanical variables where over 15% of individual samples were classified as outliers (Cook’s distance over 4 / N, where N is the number of samples [4]). Spatial sensitivity analysis compared the PLSR model performance between three adjacent NIRS measurement sites. Variables where the median coefficient of determination over the three sites was below 50% were rejected. Finally, PLSR models were gener- ated for the remaining variables using five-fold cross-validation. Models where the median cross-validated Pearson correlation coefficient was below 0.30 were rejected.

4. Results

Out of the 49 variables, 10 were selected for final analysis. The variables rejected during the elimination process were: 25 in LOO analysis, 9 in spatial sensitivity and 5 in cross-validation. Final PLSR models with calibration and cross-validation parameters are listed in Table1.

Table 1. Correlation and root-mean-square error (standardized to the range of the dependent variable) for calibration and validation of the final NIR-PLSR models.

Components rcalibration RMSEcalibration rCV RMSECV Variable

peak to equilibrium stress ratio at 4% strain 2 0.92 0.17 0.60 0.33 [0.23 - 0.43]

peak stress at 6% strain 2 0.89 0.30 0.56 0.34 [0.32 - 0.64]

equilibrium stress at 6% strain 2 0.89 0.29 0.56 0.34 [0.31 - 0.47]

peak to equilibrium stress ratio at 2% strain 2 0.90 0.23 0.56 0.32 [0.21 - 0.50]

equilibrium stress at 4% strain 2 0.89 0.26 0.55 0.32 [0.30 - 0.43]

peak stress at 4% strain 2 0.89 0.30 0.53 0.33 [0.29 - 0.69]

equilibrium stress at 2% strain 2 0.87 0.26 0.51 0.32 [0.29 - 0.58]

peak to equilibrium stress ratio at 8% strain 1 0.75 0.18 0.51 0.39 [0.25 - 0.57]

peak to equilibrium stress ratio at 6% strain 3 0.97 0.16 0.48 0.33 [0.26 - 0.54]

hysteresis percentage at 0.5 Hz 1 0.73 0.17 0.43 0.53 [0.32 - 0.66]

Nine variables of the final models were obtained from the stress-relaxation part of the mechanical testing protocol.

Models describing the ratio between peak and equilibrium stresses were retained for all four stress-relaxation steps (strains 2% - 8%), making it the most robust property for NIRS-based evaluation. Additionally, a single variable from the sinusoidal testing part corresponding to the hysteresis percentage at 0.50 Hz was also retained. Both peak- equilibrium stress ratios and sinusoidal hysteresis percentages relate to the viscous energy dissipation capabilities of the tissue. These viscoelastic properties change during a prolonged cyclic loading-unloading and can, therefore, act as indicators of decay or ultimate failure in the tissue [5]. Sensitivity towards the evaluation of these properties suggest that NIRS could be a viable diagnostics tool for determining ligament (and tendon) condition during arthroscopy.

References

1. G. Spahn, H. Plettenberg, E. Kahl, H. M. Klinger, T. M¨uckley, and G. O. Hofmann, “Near-infrared (nir) spec- troscopy. a new method for arthroscopic evaluation of low grade degenerated cartilage lesions. results of a pilot study,” BMC musculoskeletal disorders8, 47 (2007).

2. J. Ala-Myllym¨aki, J. T. Honkanen, J. T¨oyr¨as, and I. O. Afara, “Optical spectroscopic determination of human meniscus composition,” Journal of Orthopaedic Research34, 270–278 (2016).

3. C. Bonifasi-Lista, S. P. Lakez, M. S. Small, and J. A. Weiss, “Viscoelastic properties of the human medial collateral ligament under longitudinal, transverse and shear loading,” Journal of Orthopaedic Research23, 67–

76 (2005).

4. K. A. Bollen and R. W. Jackman, “Regression diagnostics: An expository treatment of outliers and influential cases,” Sociological Methods & Research13, 510–542 (1985).

5. V. C. Mow, A. Ratcliffe, and S. L. Woo, Biomechanics of diarthrodial joints, vol. 1 (Springer Science & Business Media, 2012).

Viittaukset

LIITTYVÄT TIEDOSTOT

In the double resonance spectroscopy measurements, a mid-infrared and a near- infrared light source were simultaneously used to excite two transitions of acetylene, with a shared

To test this hypothesis, we estimated the biomechanical properties (instantaneous and equilibrium moduli) of healthy and naturally degenerated bovine patellar cartilage and

However, in the present study 4 time-points (1, 2, 4 and 6 weeks post-injury) were considered, with assessment of overall tissue integrity (Mankin score) and GAG contents, thus

(d) Schematic representation of the three stage mechanical testing protocol used to obtain the 22 reference variables. Preconditioning has been omitted from the diagram... Figure 2:

Near infrared spectroscopy enables quantitative evaluation of human cartilage biomechanical properties during arthroscopy.. As a service to our customers we are providing this

In study IV, an extensive set of NIRS measurements and associated reference properties of equine articular cartilage were released in an open data publication to further facilitate

Abstract: In this study, mechanical properties of bovine knee ligament and tendon samples were estimated using near infrared spectroscopy (NIRS).. Properties related to

The same phenomena was shown, but not mentioned earlier (in Fig. 57 the weight of LV is plotted against the measurement wavelength. It can be seen that in the first LV the absorption