Altogether 77 % of the participants reported the expanded uncertainties (k=2) with their results
for at least some of their results (Table 2, Appendices 7 and 14). The range of the reported
uncertainties varied between the measurands.
Within the optimal measuring range, the expanded measurement uncertainty (k=2) should typically be 20–40%. Close to the limit of quantification the relative measurement uncertainty is higher. The harmonization of the uncertainties estimation should be continued.
Table 2. The range of the expanded measurement uncertainties (k=2, U
i%) reported by the participants.
Measurand As Ba Cl
-Conductivity Cr Cu
U
i% 12.2 – 60 % 11.9 – 58 % 10 – 62 % 8 – 33 % 13.6 – 62 % 13.1 – 60 %
Measurand DOC F
-Mo Ni pH Sb
U
i% 16 – 62 % 15 – 74 % 14.5 – 62 % 14.9 – 59 % 2.4 – 15 % 10.5 – 58 %
Measurand Se SO
42-TDS V
U
i% 18 – 61 % 10 – 62 % 14 – 65 % 15 – 58 %