: Examples of measurement uncertainties reported by the participants

In figures, the presented expanded measurement uncertainties are grouped according to the method of estimation at 95 % confidence level (k=2). The expanded uncertainties were estimated mainly by using the internal quality control (IQC) data. The used procedures in figures below are distinguished e.g. between using or not using the MUkit software for uncertainty estimation

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand 1,2,4-Trichlorobenzene Sample A1V

0 10 20 30 40

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand 1,2-Dichlorobenzene Sample N2V

0 10 20 30 40

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand 1,2-Dichloroethane Sample S3V

0

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand 1,4-Dichlorobenzene Sample A1V

0 10 20 30 40

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand Benzene Sample N2V

0 10 20 30 40

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand Chloroform Sample S3V

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand cis-1,2-Dichloroethene Sample A1V

0 10 20 30 40

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand Dichloromethane Sample N2V

0

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand ETBE Sample S3V

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand Ethylbenzene Sample A1V

0

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand m/p-Xylene Sample N2V

0 10 20 30 40

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand MTBE Sample S3V

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand o-Xylene Sample A1V

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand Styrene Sample N2V

0 10 20 30 40

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand TAME Sample S3V

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand Tetrachloroethene Sample A1V

0

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand Tetrachloromethane Sample N2V

0 10 20 30 40

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand Toluene Sample S3V

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand trans-1,2-Dichloroethene Sample A1V

0

Uncertainty, Ui %

IQC data only from synthetic control sample and/or CRM (X chart), MUkit software.

IQC data only from synthetic control sample and/or CRM (X chart), no MUkit software.

IQC data and the results obtained in proficiency tests, no MUkit software.

Data obtained from method validation, no MUkit software.

Measurand Trichloroethene Sample N2V

INTERLABORATORY PROFICIENCY TEST 15/2020

ISBN 978-952-11-5381-5 (pbk.)

FINNISH ENVIRONMENT INSTITUTE

SYKE