• Ei tuloksia

In this thesis, the current state-of-the-art MPI reconstruction and compensation methods were studied and further optimised. A special focus was on approaches which could reduce the scanning time. The possible interference of different compensation methods was evaluated and new approaches to avoid the artefacts generated by the compensation methods were devised.

The main observations of this thesis can be summarised as follows:

 Half-time MPI with attenuation and MC-based scatter correction produced images of superior quality when compared to full time imaging with standard reconstruction. One hundred thousand simulated photons in the MC-based scatter correction did not add noise to the reconstructed images even with half scanning time, but provided a good quality correction.

 Two scatter update iterations and 105 simulated photons for the 99mTc and 201Tl reconstructions and 106 simulated down-scatter photons were sufficient to obtain good quality images in the dual 99mTc/201Tl reconstruction. In order to fully compensate for the down-scatter effect in the reconstructed images, lead X-ray emissions need to be taken into account.

 The traditional reconstruction-reprojection-based motion compensation method, with three update iterations and mutual information cost function, is a good option for rapid motion correction in clinical myocardial perfusion SPECT.

 All the three Bayesian reconstruction methods studied reduced the collimator correction artefacts.

The Bowsher prior achieved a reduction without

72

evoking adverse effects on reconstructed resolution or on the contrast.

 The MPI technique can benefit greatly by incorporation of modern correction algorithms.

73

Bibliography

1. Mustajoki P. Lääkärikirja duodecim: Sepelvaltimotauti. Duodecim ; 2013.

2. Martin-Comin J, Conesa J, Pradas A. Why should you ask for a myocardial perfusion SPECT. Masson, Barcelona,2002.

3. Gibbons RS. American society of nuclear cardiology project on myocardial perfusion imaging: Measuring outcomes in response to emerging

guidelines. J Nucl Cardiol. 1996;3:436-42.

4. Underwood SR, Godman B, Salyani S, Ogle JR, Ell PJ. Economics of myocardial perfusion imaging in europe--the EMPIRE study. Eur Heart J.

1999;20:157-66.

5. Shaw LJ, Hachamovitch R, Berman DS, Marwick TH, Lauer MS, Heller GV, et al. The economic consequences of available diagnostic and prognostic strategies for the evaluation of stable angina patients: An observational assessment of the value of precatheterization ischemia.

economics of noninvasive diagnosis (END) multicenter study group. J Am Coll Cardiol. 1999;33:661-9.

6. Dvorak RA, Brown RK, Corbett JR. Interpretation of SPECT/CT myocardial perfusion images: Common artifacts and quality control techniques. Radiographics. 2011;31:2041-57.

7. El Fakhri G, Buvat I, Benali H, Todd-Pokropek A, Di Paola R. Relative impact of scatter, collimator response, attenuation, and finite spatial resolution corrections in cardiac SPECT. J Nucl Med. 2000;41:1400-8.

8. McQuaid SJ, Hutton BF. Sources of attenuation-correction artefacts in cardiac PET/CT and SPECT/CT. Eur J Nucl Med Mol Imaging.

2008;35:1117-23.

9. Narayanan MV, King MA, Pretorius PH, Dahlberg ST, Spencer F, Simon E, et al. Human-observer receiver-operating-characteristic evaluation of attenuation, scatter, and resolution compensation strategies for (99m)Tc myocardial perfusion imaging. J Nucl Med. 2003;44:1725-34.

10. Niu X, Yang Y, Jin M, Wernick MN, King MA. Effects of motion, attenuation, and scatter corrections on gated cardiac SPECT reconstruction. Med Phys. 2011;38:6571-84.

11. Zaidi H, Koral KF. Scatter modelling and compensation in emission tomography. Eur J Nucl Med Mol Imaging. 2004;31:761-82.

12. Matsumoto N, Berman DS, Kavanagh PB, Gerlach J, Hayes SW, Lewin HC, et al. Quantitative assessment of motion artifacts and validation of a new motion-correction program for myocardial perfusion SPECT. J Nucl Med.

2001;42:687-94.

13. Frey EC, Gilland KL, Tsui BM. Application of task-based measures of image quality to optimization and evaluation of three-dimensional

74

reconstruction-based compensation methods in myocardial perfusion SPECT. IEEE Trans Med Imaging. 2002;21:1040-50.

14. He X, Links JM, Gilland KL, Tsui BM, Frey EC. Comparison of 180 degrees and 360 degrees acquisition for myocardial perfusion SPECT with

compensation for attenuation, detector response, and scatter: Monte carlo and mathematical observer results. J Nucl Cardiol. 2006;13:345-53.

15. Zeintl J, Vija AH, Yahil A, Hornegger J, Kuwert T. Quantitative accuracy of clinical 99mTc SPECT/CT using ordered-subset expectation

maximization with 3-dimensional resolution recovery, attenuation, and scatter correction. J Nucl Med. 2010;51:921-8.

16. Aldridge MD, Waddington WW, Dickson JC, Prakash V, Ell PJ, Bomanji JB. Clinical evaluation of reducing acquisition time on single-photon emission computed tomography image quality using proprietary resolution recovery software. Nucl Med Commun. 2013;34:1116-23.

17. Ali I, Ruddy TD, Almgrahi A, Anstett FG, Wells RG. Half-time SPECT myocardial perfusion imaging with attenuation correction. J Nucl Med.

2009;50:554-62.

18. Armstrong IS, Arumugam P, James JM, Tonge CM, Lawson RS. Reduced-count myocardial perfusion SPECT with resolution recovery. Nucl Med Commun. 2012;33:121-9.

19. DePuey EG, Bommireddipalli S, Clark J, Leykekhman A, Thompson LB, Friedman M. A comparison of the image quality of full-time myocardial perfusion SPECT vs wide beam reconstruction half-time and half-dose SPECT. J Nucl Cardiol. 2011;18:273-80.

20. DePuey EG, Gadiraju R, Clark J, Thompson L, Anstett F, Shwartz SC.

Ordered subset expectation maximization and wide beam reconstruction

"half-time" gated myocardial perfusion SPECT functional imaging: A comparison to "full-time" filtered backprojection. J Nucl Cardiol.

2008;15:547-63.

21. Druz RS, Phillips LM, Chugkowski M, Boutis L, Rutkin B, Katz S. Wide-beam reconstruction half-time SPECT improves diagnostic certainty and preserves normalcy and accuracy: A quantitative perfusion analysis. J Nucl Cardiol. 2011;18:52-61.

22. Valenta I, Treyer V, Husmann L, Gaemperli O, Schindler MJ, Herzog BA, et al. New reconstruction algorithm allows shortened acquisition time for myocardial perfusion SPECT. Eur J Nucl Med Mol Imaging. 2010;37:750-7.

23. DePuey EG, Bommireddipalli S, Clark J, Thompson L, Srour Y. Wide beam reconstruction "quarter-time" gated myocardial perfusion SPECT functional imaging: A comparison to "full-time" ordered subset expectation maximum. J Nucl Cardiol. 2009;16:736-52.

24. Groch MW, Erwin WD. SPECT in the year 2000: Basic principles. J Nucl Med Technol. 2000;28:233-44.

25. Zaidi H. Quantitative analysis in nuclear medicine imaging. Springer, New York, 2006.

26. Hudson HM, Larkin RS. Accelerated image reconstruction using ordered subsets of projection data. IEEE Trans Med Imaging. 1994;13:601-9.

75

27. Lange K, Carson R. EM reconstruction algorithms for emission and transmission tomography. J Comput Assist Tomogr. 1984;8:306-16.

28. Shepp LA, Vardi Y. Maximum likelihood reconstruction for emission tomography. IEEE Trans Med Imaging. 1982;1:113-22.

29. Beekman FJ, Slijpen ET, Niessen WJ. Selection of task-dependent diffusion filters for the post-processing of SPECT images. Phys Med Biol.

1998;43:1713-30.

30. Hebert T, Leahy R. A generalized EM algorithm for 3-D bayesian reconstruction from poisson data using gibbs priors. IEEE Trans Med Imaging. 1989;8:194-202.

31. Nuyts J, Fessler JA. A penalized-likelihood image reconstruction method for emission tomography, compared to postsmoothed maximum-likelihood with matched spatial resolution. IEEE Trans Med Imaging.

2003;22:1042-52.

32. Seret A. Median root prior and ordered subsets in bayesian image reconstruction of single-photon emission tomography. Eur J Nucl Med.

1998;25:215-9.

33. Sohlberg A, Ruotsalainen U, Watabe H, Iida H, Kuikka JT. Accelerated median root prior reconstruction for pinhole single-photon emission tomography (SPET). Phys Med Biol. 2003;48:1957-69.

34. Bowsher JE, Johnson VE, Turkington TG, Jaszczak RJ, Floyd CR, Coleman RE. Bayesian reconstruction and use of anatomical a priori information for emission tomography. IEEE Trans Med Imaging. 1996;15:673-86.

35. Geman S, Geman D. Stochastic relaxation, gibbs distributions, and the bayesian restoration of images. IEEE Trans Pattern Anal Mach Intell.

1984;6:721-41.

36. Green PJ. Bayesian reconstructions from emission tomography data using a modified EM algorithm. IEEE Trans Med Imaging. 1990;9:84-93.

37. Alenius S. On noise reduction in iterative image reconstruction algorithms for emission tomography: Median root prior. PhD thesis, Tampere University of Technology,1999.

38. McCarthy AW, Miller MI. Maximum likelihood SPECT in clinical computation times using mesh-connected parallel computers. IEEE Trans Med Imaging. 1991;10:426-36.

39. Fleming JS. A technique for using CT images in attenuation correction and quantification in SPECT. Nucl Med Commun. 1989;10:83-97.

40. King MA, Tsui BM, Pan TS. Attenuation compensation for cardiac single-photon emission computed tomographic imaging: Part 1. impact of attenuation and methods of estimating attenuation maps. J Nucl Cardiol.

1995;2:513-24.

41. Bocher M, Balan A, Krausz Y, Shrem Y, Lonn A, Wilk M, et al. Gamma camera-mounted anatomical X-ray tomography: Technology, system characteristics and first images. Eur J Nucl Med. 2000;27:619-27.

42. Flotats A, Knuuti J, Gutberlet M, Marcassa C, Bengel FM, Kaufmann PA, et al. Hybrid cardiac imaging: SPECT/CT and PET/CT. A joint position statement by the european association of nuclear medicine (EANM), the

76

european society of cardiac radiology (ESCR) and the european council of nuclear cardiology (ECNC). Eur J Nucl Med Mol Imaging. 2011;38:201-12.

43. Ogawa K. Simulation study of triple-energy-window scatter correction in combined 201Tl, 99mTc SPECT. Ann Nucl Med. 1994;8:277-81.

44. Hashimoto J, Matsuda M, Sammiya T, Nakamura K, Tsukatani Y, Kubo A, et al. Exercise and rest myocardial scintigraphy with 201TlCl/99mTc-MIBI dual energy acquisition using triple-energy window scatter correction.

Kaku Igaku. 1994;31:1365-72.

45. Sohlberg A, Watabe H, Iida H. Acceleration of monte carlo-based scatter compensation for cardiac SPECT. Phys Med Biol. 2008;53:N277-85.

46. Beekman FJ, Kamphuis C, Frey EC. Scatter compensation methods in 3D iterative SPECT reconstruction: A simulation study. Phys Med Biol.

1997;42:1619-32.

47. Beekman FJ, Kamphuis C, King MA, van Rijk PP, Viergever MA.

Improvement of image resolution and quantitative accuracy in clinical single photon emission computed tomography. Comput Med Imaging Graph. 2001;25:135-46.

48. Beekman FJ, de Jong HW, van Geloven S. Efficient fully 3-D iterative SPECT reconstruction with monte carlo-based scatter compensation. IEEE Trans Med Imaging. 2002;21:867-77.

49. Bokulic T, Vastenhouw B, de Jong HW, van Dongen AJ, van Rijk PP, Beekman FJ. Monte carlo-based down-scatter correction of SPECT attenuation maps. Eur J Nucl Med Mol Imaging. 2004;31:1173-81.

50. Kamphuis C, Beekman FJ, van Rijk PP, Viergever MA. Dual matrix ordered subsets reconstruction for accelerated 3D scatter compensation in single-photon emission tomography. Eur J Nucl Med. 1998;25:8-18.

51. Sohlberg A, Watabe H, Iida H. Three-dimensional SPECT reconstruction with transmission-dependent scatter correction. Ann Nucl Med.

2008;22:549-56.

52. Frey EC, Tsui BM, Gullberg GT. Improved estimation of the detector response function for converging beam collimators. Phys Med Biol.

1998;43:941-50.

53. Song X, Segars WP, Du Y, Tsui BM, Frey EC. Fast modelling of the collimator-detector response in monte carlo simulation of SPECT imaging using the angular response function. Phys Med Biol. 2005;50:1791-804.

54. Metz CE, Atkins FB, Beck RN. The geometric transfer function component for scintillation camera collimators with straight parallel holes. Phys Med Biol. 1980;25:1059-70.

55. Du Y, Frey EC, Wang WT, et al. Combination of MCNP and SimSET for monte carlo simulation of SPECT with medium- and high-energy photons.

IEEE Trans Nucl Science. 2002;49:668-74.

56. Wilderman SJ, Dewaraja Y, Koral KF. Accurate modeling of nuclear-medicine collimators in monte carlo simulation of high energy photons.

Nuclear Instruments and Methods in Physics Research. 1999;A 422:745-50.

77

57. Tsui BM, Hu HB, Gilland DR, et al. Implementation of simultaneous attenuation and detector response correction in SPECT. IEEE Trans Nucl Science. 1988;35:778-83.

58. Qi ZZ, Tian R. Impact of patient motion on myocardial perfusion imaging with double-head SPECT and assessment of motion correction. Sichuan Da Xue Xue Bao Yi Xue Ban. 2011;42:392-6.

59. Massardo T, Jaimovich R, Faure R, Munoz M, Alay R, Gatica H. Motion correction and myocardial perfusion SPECT using manufacturer provided software. does it affect image interpretation? Eur J Nucl Med Mol Imaging.

2010;37:758-64.

60. Germano G, Chua T, Kavanagh PB, Kiat H, Berman DS. Detection and correction of patient motion in dynamic and static myocardial SPECT using a multi-detector camera. J Nucl Med. 1993;34:1349-55.

61. Feng B, Gifford HC, Beach RD, Boening G, Gennert MA, King MA. Use of three-dimensional gaussian interpolation in the projector/backprojector pair of iterative reconstruction for compensation of known rigid-body motion in SPECT. IEEE Trans Med Imaging. 2006;25:838-44.

62. Mukherjee JM, Hutton BF, Johnson KL, Pretorius PH, King MA. An evaluation of data-driven motion estimation in comparison to the usage of external-surrogates in cardiac SPECT imaging. Phys Med Biol.

2013;58:7625-46.

63. Bai C, Maddahi J, Kindem J, Conwell R, Gurley M, Old R. Development and evaluation of a new fully automatic motion detection and correction technique in cardiac SPECT imaging. J Nucl Cardiol. 2009;16:580-9.

64. Zaret B, Beller G. Clinical nuclear cardiology: State of the art and future directions. Mosby, Philadelphia, 2010.

65. Hesse B, Tagil K, Cuocolo A, Anagnostopoulos C, Bardies M, Bax J, et al.

EANM/ESC procedural guidelines for myocardial perfusion imaging in nuclear cardiology. Eur J Nucl Med Mol Imaging. 2005;32:855-97.

66. Tsui BM, Frey EC, LaCroix KJ, Lalush DS, McCartney WH, King MA, et al.

Quantitative myocardial perfusion SPECT. J Nucl Cardiol. 1998;5:507-22.

67. Hendel RC, Corbett JR, Cullom SJ, DePuey EG, Garcia EV, Bateman TM.

The value and practice of attenuation correction for myocardial perfusion SPECT imaging: A joint position statement from the american society of nuclear cardiology and the society of nuclear medicine. J Nucl Cardiol.

2002;9:135-43.

68. O'connor MK, Kemp B, Anstett F, Christian P, Ficaro EP, Frey E, et al. A multicenter evaluation of commercial attenuation compensation techniques in cardiac SPECT using phantom models. J Nucl Cardiol.

2002;9:361-76.

69. King MA, Tsui BM, Pan TS, Glick SJ, Soares EJ. Attenuation compensation for cardiac single-photon emission computed tomographic imaging: Part 2.

attenuation compensation algorithms. J Nucl Cardiol. 1996;3:55-64.

70. Enevoldsen LH, Menashi CA, Andersen UB, Jensen LT, Henriksen OM.

Effects of acquisition time and reconstruction algorithm on image quality,

78

quantitative parameters, and clinical interpretation of myocardial perfusion imaging. J Nucl Cardiol. 2013;20:1086-92.

71. McNamara JE, Pretorius PH, Johnson K, Mukherjee JM, Dey J, Gennert MA, et al. A flexible multicamera visual-tracking system for detecting and correcting motion-induced artifacts in cardiac SPECT slices. Med Phys.

2009;36:1913-23.

72. Veress AI, Segars WP, Weiss JA, Tsui BM, Gullberg GT. Normal and pathological NCAT image and phantom data based on physiologically realistic left ventricle finite-element models. IEEE Trans Med Imaging.

2006;25:1604-16.

73. Ljungberg M, Strand SE. A monte carlo program for the simulation of scintillation camera characteristics. Comput Methods Programs Biomed.

1989;29:257-72.

74. Lowe VJ, Greer KL, Hanson MW, Jaszczak RJ, Coleman RE. Cardiac phantom evaluation of simultaneously acquired dual-isotope rest thallium-201/stress technetium-99m SPECT images. J Nucl Med.

1993;34:1998-2006.

75. Hannequin P, Mas J, Germano G. Photon energy recovery for crosstalk correction in simultaneous 99mTc/201Tl imaging. J Nucl Med.

2000;41:728-36.

76. de Jong HW, Beekman FJ, Viergever MA, van Rijk PP. Simultaneous (99m)Tc/(201)Tl dual-isotope SPET with monte carlo-based down-scatter correction. Eur J Nucl Med Mol Imaging. 2002;29:1063-71.

77. Weinmann P, Foult JM, Le Guludec D, Tamgac F, Rechtman D, Neuman A, et al. Dual-isotope myocardial imaging: Feasibility, advantages and limitations. preliminary report on 231 consecutive patients. Eur J Nucl Med. 1994;21:212-5.

78. Kadrmas DJ, Frey EC, Tsui BM. Simultaneous technetium-99m/thallium-201 SPECT imaging with model-based compensation for

cross-contaminating effects. Phys Med Biol. 1999;44:1843-60.

79. Hademenos GJ, Dahlbom M, Hoffman EJ. Simultaneous dual-isotope technetium-99m/thallium-201 cardiac SPET imaging using a projection-dependent spilldown correction factor. Eur J Nucl Med. 1995;22:465-72.

80. Kiat H, Germano G, Friedman J, Van Train K, Silagan G, Wang FP, et al.

Comparative feasibility of separate or simultaneous rest thallium-201/stress technetium-99m-sestamibi dual-isotope myocardial perfusion SPECT. J Nucl Med. 1994;35:542-8.

81. Nakamura M, Takeda K, Ichihara T, Motomura N, Shimizu H, Saito Y, et al. Feasibility of simultaneous stress 99mTc-sestamibi/rest 201Tl dual-isotope myocardial perfusion SPECT in the detection of coronary artery disease. J Nucl Med. 1999;40:895-903.

82. Matsunari I, Boning G, Ziegler SI, Kosa I, Nekolla SG, Ficaro EP, et al.

Attenuation-corrected rest thallium-201/stress technetium-99m sestamibi myocardial SPECT in normals. J Nucl Cardiol. 1998;5:48-55.

83. Unlu M, Gunaydin S, Ilgin N, Inanir S, Gokcora N, Gokgoz L. Dual isotope myocardial perfusion SPECT in the detection of coronary artery

79

disease: Comparison of separate and simultaneous acquisition protocols. J Nucl Biol Med. 1993;37:233-7.

84. Berman DS, Kiat H, Friedman JD, Wang FP, van Train K, Matzer L, et al.

Separate acquisition rest thallium-201/stress technetium-99m sestamibi dual-isotope myocardial perfusion single-photon emission computed tomography: A clinical validation study. J Am Coll Cardiol. 1993;22:1455-64.

85. Dey T, Backus BE, Romijn RL, Wieczorek H, Verzijlbergen JF. Low-dose single acquisition rest Tc/stress Tl myocardial perfusion SPECT protocol:

Phantom studies and clinical validation. Eur J Nucl Med Mol Imaging.

2013.

86. Sohlberg AO, Kajaste MT. Fast monte carlo-simulator with full collimator and detector response modelling for SPECT. Ann Nucl Med. 2012;26:92-8.

87. Slomka PJ, Nishina H, Berman DS, Kang X, Friedman JD, Hayes SW, et al.

Automatic quantification of myocardial perfusion stress-rest change: A new measure of ischemia. J Nucl Med. 2004;45:183-91.

88. Prasad M, Slomka PJ, Fish M, Kavanagh P, Gerlach J, Hayes S, et al.

Improved quantification and normal limits for myocardial perfusion stress-rest change. J Nucl Med. 2010;51:204-9.

89. Liu S, Farncombe T. IEEE nuclear science symposium conference record. . 2007;6:3955.

90. Snyder D, Miller M, Thomas L, Politte D. Noise and edge artefacts in maximum likelihood reconstruction for emission tomography. 1987;6:228.

91. Alenius S, Ruotsalainen U. Bayesian image reconstruction for emission tomography based on median root prior. Eur J Nucl Med. 1997;24:258-65.

92. Lehovich A, Gifford HC, Schneider PB, King MA. Choosing anatomical-prior strength for MAP SPECT reconstruction to maximize lesion detectability. IEEE Nucl Sci Symp Conf Rec (1997). 2007;6:4222-5.

93. Gindi G, Lee M, Rangarajan A, Zubal IG. Bayesian reconstruction of functional images using anatomical information as priors. IEEE Trans Med Imaging. 1993;12:670-80.

94. Bateman TM, Heller GV, McGhie AI, Courter SA, Golub RA, Case JA, et al.

Multicenter investigation comparing a highly efficient half-time stress-only attenuation correction approach against standard rest-stress Tc-99m SPECT imaging. J Nucl Cardiol. 2009;16:726-35.

95. Borges-Neto S, Pagnanelli RA, Shaw LK, Honeycutt E, Shwartz SC, Adams GL, et al. Clinical results of a novel wide beam reconstruction method for shortening scan time of Tc-99m cardiac SPECT perfusion studies. J Nucl Cardiol. 2007;14:555-65.

96. Venero CV, Heller GV, Bateman TM, McGhie AI, Ahlberg AW, Katten D, et al. A multicenter evaluation of a new post-processing method with depth-dependent collimator resolution applied to full-time and half-time acquisitions without and with simultaneously acquired attenuation correction. J Nucl Cardiol. 2009;16:714-25.

97. Takahashi Y, Murase K, Mochizuki T, Sugawara Y, Maeda H, Kinda A.

Simultaneous 3-dimensional resolution correction in SPECT

80

reconstruction with an ordered-subsets expectation maximization algorithm. J Nucl Med Technol. 2007;35:34-8.

98. de Wit T, Jianbin X, Beekman F. Monte carlo-based statistical SPECT reconstruction: Influence of number of photon tracks. 2005;52:1365.

99. Lau YH, Hutton BF, Beekman FJ. Choice of collimator for cardiac SPET when resolution compensation is included in iterative reconstruction. Eur J Nucl Med. 2001;28:39-47.

100. Groch MW, Takamiya Y, Groch PJ, Erwin WD. Quantitative gated myocardial SPECT: Effect of collimation on left-ventricular ejection fraction. J Nucl Med Technol. 2000;28:36-40.

101.Zhang B, Zeng GL. High-resolution versus high-sensitivity SPECT imaging with geometric blurring compensation for various parallel-hole collimation geometries. IEEE Trans Inf Technol Biomed. 2010;14:1121-7.

102. Hofheinz F,Dittrich S, Pötzsch C, Hoff Jv. Effects of cold sphere walls in PET phantom measurements on the volume reproducing threshold. Phys Med Biol. 2010;55:1099-113.

103. Nakazato R, Berman DS, Hayes SW, Fish M, Padgett R, Xu Y, Lemley M, Baavour R, Roth N, Slomka PJ. Myocardial perfusion imaging with a solid-state camera: simulation of a very low dose imaging protocol. J Nucl Med. 2013;54:373-9.

104.Verger A, Djaballah W, Fourquet N, Rouzet F, Koehl G, Imbert L, Poussier S, Fay R, Roch V, Le Guludec D, Karcher G, Marie PY. Comparison between stress myocardial perfusion SPECT recorded with cadmium-zinc-telluride and Anger cameras in various study protocols. Eur J Nucl Med Mol Imaging. 2013;40:331-40.

105.Gimelli A, Bottai M, Genovesi D, Giorgetti A, Di Martino F, Marzullo P.

High diagnostic accuracy of low-dose gated-SPECT with solid-state ultrafast detectors: preliminary clinical results. Eur J Nucl Med Mol Imaging. 2012;39:83-90.

106.Smith MF. Recent advances in cardiac SPECT instrumentation and system design. Curr Cardiol Rep. 2013;15:387.

Publications of the University of Eastern Finland Dissertations in Forestry and Natural Sciences

Publications of the University of Eastern Finland Dissertations in Forestry and Natural Sciences

isbn 978-952-61-1505-4

Tuija Kangasmaa

Evaluation and Optimization

of Novel Reconstruction Methods for Myocardial Perfusion SPECT

Myocardial perfusion imaging (MPI) is one of the most common types of SPECT studies, being highly valued for its diagnostic accuracy. However, image noise, photon attenuation, Compton scatter, collimator-detector response (CDR) and patient motion hamper the image quality of MPI. These image degrading factors have been investigated widely and the modern reconstruction-based compensation methods can greatly improve the image quality. Recently CDR compensation has attracted considerable interest, because it allows diagnostically satisfying images to be acquired in half of the acquisition time currently in use.

Shorter scan times both reduce artifacts related to patient motion and increase the patient throughput. The aim of this thesis was to validate and optimize novel SPECT reconstruction and compensation techniques for use with MPI. The specific focus was on scan time reduction and on CDR compensation.

dissertations | 142 | Tuija Kangasmaa | Evaluation and Optimization of Novel Reconstruction Methods for Myocardial Perfusion SPECT

dissertations | 142 | Tuija Kangasmaa | Evaluation and Optimization of Novel Reconstruction Methods for Myocardial Perfusion SPECT